Please use this identifier to cite or link to this item:
https://ric.cps.sp.gov.br/handle/123456789/29882
Title: | Validação de metodologia analítica de cromatografia líquida de alta eficiência para controle de qualidade de radiofármaco 18F PSMA-1007 |
Authors: | GODOY, Rodrigo |
Advisor: | CORBI, Fabiana Cristina Andrade |
Other contributor: | CORBI, Fabiana Cristina Andrade REIS, Aurimar Moreira BRANCO, Luciene Maria G. Castello |
type of document: | Monografia |
Keywords: | Radiofarmacologia;Cromatografia líquida |
Issue Date: | 23-Jun-2023 |
Publisher: | 276 |
Citation: | GODOY, Rodrigo. Validação de metodologia analítica de cromatografia liquida de alta eficiência para controle de qualidade de radiofármaco 18F PSMA-1007 |
Abstract: | RESUMO
A validação da metodologia analítica de cromatografia líquida de alta eficiência (HPLC) para
o controle de qualidade do radiofármaco 18F-PSMA-1007 foi realizada utilizando um processo
essencial, que envolve a determinação de parâmetros de desempenho, estudos de robustez e
estabilidade, seletividade, linearidade, precisão, exatidão e limites de detecção e quantificação.
A partir dos parâmetros utilizados, foi possível garantir a segurança e a eficácia dos
radiofármacos que são utilizados em exames de imagem molecular de próstata, além da
metodologia atender os requisitos obrigatórios da norma regulamentadora RDC 166. Os
resultados demonstraram que o método empregado é eficaz para a determinação da impureza,
mas não apresentou um grau de robustez confiável, o que não inviabiliza a utilização deste
método, mas obriga o acompanhamento dos parâmetros definidos na robustez que não tiveram
um desempenho satisfatório, como a variação da concentração do gradiente e a coluna de
fabricantes diferentes. Para os demais parâmetros exigidos, o método demonstrou um alto grau
de confiabilidade.
Palavras-chave: validação; radiofármaco; metodologia. ABSTRACT The validation of the analytical methodology of high performance liquid chromatography (HPLC) for the quality control of the radiopharmaceutical 18F-PSMA-1007 was carried out using an essential process, which involves the excellence of performance standards, studies of robustness and stability, selectivity, linearity, precision, accuracy and limits of detection and quantification. From the parameters used, it was possible to guarantee the safety and efficacy of the radiopharmaceuticals that are used in molecular imaging tests of the prostate, the methodology meeting the mandatory requirements of the regulatory standard RDC 166. The results proved that the applied method is also effective for the impurity projected, but did not present a reliable degree of robustness, which does not make the use of this method unfeasible, but requires the monitoring of the parameters defined in the robustness that did not have a projected performance, such as the variation in the concentration of the gradient and the column of different manufacturers. For the other required parameters, the method demonstrated a high degree of reliability. Keywords: validation; radiopharmaceutical; methodology. |
URI: | https://ric.cps.sp.gov.br/handle/123456789/29882 |
Appears in Collections: | Trabalhos de Conclusão de Curso |
Files in This Item:
File | Description | Size | Format | |
---|---|---|---|---|
Tecnologiaemprocessosquimicos_1_2023_Rodrigo Godoy_Validação de metodologia analitica de cromatografia liquida de alta eficiencia para controle de qualidade de ....pdf | 1.45 MB | Adobe PDF | View/Open |
Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.